LC-MS Worklist Builder
io.github.K-Dense-AIPublished by GitHub account K-Dense-AIv0.1.1Updated Oct 25, 2026
MCP server that builds and validates LC-MS worklists for MassLynx, SCIEX OS, MassHunter, Xcalibur.
Overview
A local MCP server that lets an assistant build, validate and export LC-MS sample worklists as import files for MassLynx, SCIEX OS, MassHunter and Xcalibur.
- What you can do
- Create and extend in-memory draft worklists with methods, injection volumes, tray positions and data-file naming patterns, and add samples to them. Insert blanks and QC injections and randomise run order with a recorded seed. Validate drafts for missing fields, duplicate data-file names, illegal Windows filename characters, position bounds, injection-volume limits and method extensions. Export the draft as a vendor import file into one output folder, with a preview, warnings and import instructions. Import an existing vendor file to validate, edit or convert it to another vendor's format, and list supported formats, positions and stored worklists.
- When to use it
- Useful when preparing LC-MS batches for Waters MassLynx, SCIEX OS, Agilent MassHunter or Thermo Xcalibur and you want an assistant to lay out samples, add blanks and QCs, randomise order and catch mistakes before a person imports the file. Also handy for converting an existing batch between vendors.
- Requirements
- Runs locally as a Python package (uvx labmcp-ms-worklist) over stdio; desktop only. Needs an output folder given by LABMCP_ADDRESS (default ./worklists), created if missing. Optional LABMCP_LIMITS overrides safety limits, LABMCP_READ_ONLY=1 hides state-changing tools, and LABMCP_SIMULATE=1 uses a built-in simulator with a temporary folder. No authentication or API keys.
Installation
In SourceWeft
- Open LC-MS Worklist Builder in the dashboard and add it to a workspace.
- Enable the server for the chats that should use its tools.
Desktop only via STDIO. STDIO servers start a local process, so they need the SourceWeft desktop host.
Other MCP clients
Follow the launch instructions in the repository.
README
LC-MS Worklist Builder (MassLynx, SCIEX OS, MassHunter, Xcalibur): MCP Server
Let an AI agent build LC-MS sample queues (worklists, sequences, batches) and export them as import files for Waters MassLynx, SCIEX OS, Agilent MassHunter Acquisition and Thermo Xcalibur. The agent can lay out samples on a tray, insert blanks and QCs, randomise the run order with a recorded seed, check the list for mistakes, convert an existing file from one vendor to another, and write the file for a person to import. It uses only documented import formats. It never starts an acquisition and doesn't talk to the mass spectrometer.
Try it without an instrument
--simulate uses a temporary output folder that is deleted when the server stops. Everything else is the same code.
Set up
The "instrument address" is the output folder where worklist files are written (default ./worklists, created if missing):
- The server only reads and writes inside this folder. It refuses
.., absolute paths elsewhere and symlinks that point outside it. It never overwrites an existing file unless the agent passesoverwrite=true. - To edit or convert an existing file, or to use a vendor template, copy it into the folder first.
- Point
--addressat a folder the acquisition software watches only if the vendor documents that behaviour for your setup. Otherwise import the file by hand (instructions below).
Add to your MCP client
Claude Code
Claude Desktop / Cursor / Windsurf
--read-only keeps only the tools that inspect things (list_formats, validate_worklist, import_worklist, get_worklist, list_worklists).
Tools
get_connection_info, get_command_log and reconnect are built into every LabMCP server. There are no HAZARD tools: nothing here moves, heats, injects or acquires. The server only writes files. The CONTROL tools change in-memory drafts or write a file, and --read-only hides them.
Typical flow: create_worklist → add_samples (optional) → insert_qc_blanks → validate_worklist → export_worklist, then import the file in the vendor software.
Formats
The table below separates what the vendor documents say (verified) from what the server assumes. list_formats returns the same information to the agent. To be sure of the header for your software version, export an empty batch, sequence or worklist from your own system, put it in the output folder, and pass it to export_worklist as template_path. The server then writes that header, and that delimiter, exactly.
Waters MassLynx: sample list worksheet (waters_masslynx)
- Import: Sample list > File > Import Worksheet, File of Type "Comma Delimited (*.CSV, *.TXT)".
- Verified:
- Row 1 holds the MassLynx FIELD IDs, which are case sensitive. Columns can be in any order, and an
Indexcolumn is added. Source: WKB63781. - The required columns are
FILE_NAME,INLET_FILE,MS_FILE,SAMPLE_LOCATIONandINJ_VOL. Source: MassLynx 4.2 Getting Started Guide 715009602, Table 5-1. The "Bottle" column's FIELD ID isSAMPLE_LOCATION, notBOTTLE. - The example files attached to WKB63781 use CRLF line endings and unquoted fields,
TYPEvaluesBlank/Standard/QC/Analyte, and method names without an extension.
- Row 1 holds the MassLynx FIELD IDs, which are case sensitive. Columns can be in any order, and an
- Mapping:
FILE_NAME= data file,FILE_TEXT= sample name,MS_FILE,MS_TUNE_FILE,INLET_FILE,SAMPLE_LOCATION,TYPE,ID,INJ_VOL,Index. You can add any other FIELD ID (for exampleCONC_AorQUAN_REF) per sample throughextra. - Assumed:
solventanddouble_blankare written asBlank.- Extension checks:
.expfor MS_FILE and.iprfor the tune file, applied only when an extension is given. - Files are UTF-8 without a BOM. Non-ASCII text is flagged.
- The
waters_plate_wellposition pattern (1:A,1) is a common convention. Its comma is CSV-quoted.
SCIEX OS: batch import (sciex_os)
- Import: Batch workspace > Open > Import from file, then Save As.
- Verified:
- Batches import from
.txtor.csv, and the column layout must come from a batch exported from SCIEX OS. Source: SCIEX KB. - Column names Sample Name (< 252 characters), MS Method, Sample Type, Data File and Processing Method. Source: SCIEX OS Feature Guide RUO-IDV-05-15796-C, Table 4-1.
- Rack Type, Rack Position, Plate Type, Plate Position and Vial Position columns. Source: SCIEX KB.
- Sample types Unknown, Blank, Standard, QualityControl, Solvent and Double blank.
- Several samples can share one data file.
- Batches import from
- Assumed:
- SCIEX doesn't publish the exact header text of an exported batch. The default header (
Sample Name,Sample ID,Sample Type,MS Method,LC Method,Rack Type,Rack Position,Plate Type,Plate Position,Vial Position,Injection Volume,Data File,Processing Method,Comment) is a best guess. Usetemplate_pathwith your own export.validate_worklistandexport_worklistwarn when no template was used. - Extensions
.msm,.lcmand.qmethodproduce warnings only. - Rack and plate columns are set with
vendor_columnsor per sample withextra.
- SCIEX doesn't publish the exact header text of an exported batch. The default header (
Agilent MassHunter Acquisition: worklist CSV import (agilent_masshunter)
- Import: Worklist, right-click > "Add/Append Samples from... (csv, xlsx)". Older versions call it "Import Worklist...". You don't need a map file because the headers match the worklist column names.
- Verified:
- The header
Sample Name,Barcode,Rack Code,Sample Position,Method,Data File,Sample Type,Level Name,Inj Vol (µL),Comment, positions likeP1-A1, and the typesCalibrationandSamplecome from the example file shipped inD:\MassHunter\Worklist_Import, as quoted on the Agilent Community. - Save as UTF-8 when the file contains µ. Source: Agilent Community.
Inj Vol=-1means "As method". Source: Agilent Known Problem Report.- Sample Position, Method and Data File must be filled in. Source: Study Manager Quick Start G3335-90140.
- The header
- Assumed:
- The file has a UTF-8 BOM and CRLF line endings.
- Data files get a
.dsuffix. - Methods should end in
.m. - The sample types Blank, QC and DoubleBlank don't appear in the example;
solventis written as Blank.
Thermo Xcalibur: sequence CSV import (thermo_xcalibur)
- Import: Sequence Setup > File > Import Sequence.
- Verified in the Xcalibur 2.2 Acquisition and Processing User Guide XCALI-97209 Rev. D, Appendix A:
- Only
.csvfiles are accepted. - The first cell must be
Bracket Type=n(1 Overlapped, 2 None, 3 Non-Overlapped, 4 Open). - The separator must match the Windows list separator. Use
delimiter=";"where that is;. - The sequence column names and the sample types Unknown, Blank, QC, Std Bracket, Std Clear and Std Update.
- Only
- Corroborated, not published by Thermo: the exact row-2 header strings (
Sample Type,File Name,Sample ID,Path,Instrument Method,Process Method,Calibration File,Position,Inj Vol,Level,Sample Wt,Sample Vol,ISTD Amt,Dil Factor,L1 Study,L2 Client,L3 Laboratory,L4 Company,L5 Phone,Comment,Sample Name). Open-source generators that target Xcalibur import use them: protti, fgcz/qg and lcms-sequencer. Native exports start with a bareBracket Type=4line (Rapid-QC-MS #83). Export one sequence from your Xcalibur as a template if you want certainty. - Assumed:
- Sample Wt 0, Sample Vol 0, ISTD Amt 0 and Dil Factor 1 are written as defaults.
standardis written asStd Bracket, which suits bracket type 4.- Extension checks:
.methfor the instrument method and.pmdfor the processing method.
Not implemented
- MassHunter Walkup Custom Sample Import (G2725-90026): the guide couldn't be retrieved (HTTP 403), so its format couldn't be verified.
- Native worklist and sequence files (MassHunter
.wklXML, Xcalibur.sld, MassLynx.spl): not implemented. We found no public specification for them, so the server uses the CSV import paths above.
Validation
validate_worklist, which export_worklist also runs, reports errors, which block export, and warnings:
- Required fields for the format are missing.
- Duplicate data-file names, compared case-insensitively. For SCIEX OS a shared data file is allowed, but a sample name repeated within one data file is an error.
- Characters Windows doesn't allow in file names (
< > : " / \ | ? *, control characters), a trailing dot or space, and reserved names (CON, NUL, COM1…). - The position doesn't match the chosen
position_pattern(vial_number,well,agilent_plate_well,waters_plate_well,tray_well) or falls outside theplate_size(24/48/54/96/384) ormax_vial. - The injection volume is ≤ 0, not a number, or above the
max_injection_volume_ullimit. This includes injection-volume columns set throughextraorvendor_columns. - A text field or
extravalue contains a line break or another control character, which would break the import file. - The method file has the wrong extension, for example an Analyst
.damin a SCIEX OS batch, or a MassHunter.min an Xcalibur sequence. - Fields the target format has no column for, and non-ASCII text in formats without a BOM.
Safety limits
Override at launch, for example --limit max_injection_volume_ul=20 for a 20 µL loop.
Example prompts
- "Make an Xcalibur sequence for the 40 plasma samples in this list: wells A1 onwards on the R tray, 2 µL, method C:\Xcalibur\methods\DIA_60min. Add a blank after every 10 samples from R:H12 and a pooled QC at the start and end from R:H11. Randomise the samples and tell me the seed."
- "Import
old_batch.csvfrom the worklist folder and convert it to a MassHunter worklist with method D:\MassHunter\methods\Panel.m." - "I put
blank_batch.txt(exported from our SCIEX OS) in the folder. Use it as the template for this batch." - "Check this MassLynx sample list for duplicate file names and bad vial positions before I import it."
Notes
- Drafts live in memory.
reconnect(or restarting the server) discards them. In--simulateit also deletes the temporary folder. Export files you want to keep. - By default,
export_worklistalso writes<file>.provenance.jsonnext to the worklist (for examplePlasma.csv.provenance.json). Like the worklist itself, it is never overwritten unlessoverwrite=true. It contains the operation history, the randomisation seed and the sample order before randomisation, so the run order can be reproduced. Passwrite_provenance=falseto skip it. blank_every_n/qc_every_nalso add a control after the last sample when the count divides evenly, unlessblank_at_end/qc_at_endalready adds one there.- When a draft is exported in another vendor's format, values kept verbatim from the original vendor for columns the new vendor maps itself (for example Xcalibur's
Std Clearsample type) are not copied; the export notes say so. - Randomisation shuffles only the types in
randomize_types(by default unknown samples) among their own slots, using Python'srandom.Random(seed). Standards, blanks and QCs stay where they are. - Method names must match methods that exist on the acquisition PC. The server can't check that.
Verification
Source: servers/chemistry/ms-worklist/README.md at commit 46545ea
Tools
0Version history
1- v0.1.1LatestSep 27, 2026
Score
50% of the applicable checks met
How complete this listing is, computed from the registry entry, the endpoint and public sources. Not a rating of the server and not a statement that it is safe or working today: reachability is a dated check, and no installs or reviews go into the score.
- Reached in the last 30 days25 pts
Keep the endpoint answering the MCP handshake; SourceWeft probes it every 30 days.
- Auth declared10 pts
Declare the credential headers in the registry entry, or keep the endpoint answering the probe so its auth can be classified.
- Tools listed15 pts
Answer tools/list without a credential, or ship a package that starts in a sandbox without one.
- DescriptionRequired15 pts
- Logo10 pts
- Documentation15 pts
- License10 pts