Take one reading with the function, range and speed currently set on the instrument.
Function, range and speed are kept; the trigger is set to immediate with one reading
per trigger, as measure does. Returns value, unit, range, timestamp (UTC) and, where
the data sheet allows, the accuracy.
Takes no parameters.
Select a measurement function and take one reading.
This switches the instrument (front panel included) to the given function, with autorange
unless a range is given. Returns value, unit, range used, timestamp (UTC) and, where the data
sheet allows, the accuracy of the reading.
Args:
function: Measurement function. Leads in HI/LO for voltage, resistance_2w, capacitance,
frequency, period, continuity, diode and temperature; in I/LO for currents; resistance_4w
also uses HI Sense/LO Sense.
range: Fixed range in V, A, Ω or F; omit for autorange. dc_voltage 0.6, 6, 60, 600, 1000;
ac_voltage 0.6, 6, 60, 600, 750; dc_current 0.0006, 0.006, 0.06, 0.6, 6, 10; ac_current
0.06, 0.6, 6, 10; resistance_2w/resistance_4w 600, 6000, 60000, 600000, 6e6, 6e7, 1e8;
capacitance 2e-9, 2e-8, 2e-7, 2e-6, 2e-5, 0.0002, 0.01. Other functions have no range.
speed: Only for dc_voltage, dc_current, resistance_2w, resistance_4w: slow (10 PLC, default,
most accurate), medium (1 PLC), fast (0.3 PLC).
sensor: Required for temperature: pt100 (RTD) or thermocouple type B, E, J, K, N, R, S or T.
Parametersrangespeedsensorfunctionrequired
Record timestamped readings at a fixed interval and return statistics plus all samples.
Readings are paced by this computer. A reading takes about 0.3 s at speed slow and in most
other functions, 0.6 s for frequency and period; with a shorter interval readings are taken
back to back, and none starts later than one interval after duration_s, so the recording
ends on time with fewer samples. Without function the present function and range are kept;
otherwise they are set as in measure.
Args:
duration_s: Recording time in seconds (max 600).
interval_s: Time between readings in seconds (min 0.1).
function: Measurement function. Leads in HI/LO for voltage, resistance_2w, capacitance,
frequency, period, continuity, diode and temperature; in I/LO for currents; resistance_4w
also uses HI Sense/LO Sense.
range: Fixed range in V, A, Ω or F; omit for autorange. dc_voltage 0.6, 6, 60, 600, 1000;
ac_voltage 0.6, 6, 60, 600, 750; dc_current 0.0006, 0.006, 0.06, 0.6, 6, 10; ac_current
0.06, 0.6, 6, 10; resistance_2w/resistance_4w 600, 6000, 60000, 600000, 6e6, 6e7, 1e8;
capacitance 2e-9, 2e-8, 2e-7, 2e-6, 2e-5, 0.0002, 0.01. Other functions have no range.
speed: Only for dc_voltage, dc_current, resistance_2w, resistance_4w: slow (10 PLC, default,
most accurate), medium (1 PLC), fast (0.3 PLC).
sensor: Required for temperature: pt100 (RTD) or thermocouple type B, E, J, K, N, R, S or T.
Parametersrangespeedsensorfunctionduration_sinterval_s
Take a series of readings as fast as the instrument can and return statistics plus all values.
Readings are paced by the instrument and have no individual timestamps. Approximate rates
(50 Hz mains): dc_voltage, dc_current and resistance 4/s at speed slow, 40/s medium, 175/s
fast; ac functions and temperature 4/s; frequency and period 2/s. Bursts estimated to take
longer than 600 s are refused. Without function the present function and range are kept;
otherwise they are set as in measure.
Args:
count: Number of readings, 1 to 10000.
function: Measurement function. Leads in HI/LO for voltage, resistance_2w, capacitance,
frequency, period, continuity, diode and temperature; in I/LO for currents; resistance_4w
also uses HI Sense/LO Sense.
range: Fixed range in V, A, Ω or F; omit for autorange. dc_voltage 0.6, 6, 60, 600, 1000;
ac_voltage 0.6, 6, 60, 600, 750; dc_current 0.0006, 0.006, 0.06, 0.6, 6, 10; ac_current
0.06, 0.6, 6, 10; resistance_2w/resistance_4w 600, 6000, 60000, 600000, 6e6, 6e7, 1e8;
capacitance 2e-9, 2e-8, 2e-7, 2e-6, 2e-5, 0.0002, 0.01. Other functions have no range.
speed: Only for dc_voltage, dc_current, resistance_2w, resistance_4w: slow (10 PLC, default,
most accurate), medium (1 PLC), fast (0.3 PLC).
sensor: Required for temperature: pt100 (RTD) or thermocouple type B, E, J, K, N, R, S or T.
Parameterscountrangespeedsensorfunction
Instrument model, serial number, firmware, VISA resource and the present measurement configuration.
Takes no parameters.